Chemiluminescence Immunoassay System

  • Chemiluminescence Immunoassay System
  • Chemiluminescence Immunoassay System
  • Chemiluminescence Immunoassay System
  • Chemiluminescence Immunoassay System
Chemiluminescence Immunoassay System
  • /
  • China
  • 30 days
  • 5000pcs/monthly

Introduction:
The chemiluminescence immunoassay system uses magnetic particle separation technology, which uses magnetic particles as antibody carriers, which can be evenly distributed in the liquid phase reaction system, with faster reaction speed and higher efficiency. Using enzymatic chemiluminescence method, the light signal is more stable. A new generation of enzymatic substrates, with higher sensitivity and faster luminescence.

Introduction:
The chemiluminescence immunoassay system uses magnetic particle separation technology, which uses magnetic particles as antibody carriers, which can be evenly distributed in the liquid phase reaction system, with faster reaction speed and higher efficiency. Using enzymatic chemiluminescence method, the light signal is more stable. A new generation of enzymatic substrates, with higher sensitivity and faster luminescence.

1. Operation Module:

Tubular enzymatic chemiluminescence

The max speed of a single machine is 240T/h

Carrying contamination rate<10-5

High precision: intra-batch precision≤8%

2. Control Module:

Ergonomic design increased operator comfort

0.68 square meters, highly integrated and simplified patented design

Convenient operation: support LIS two-way communication, scan the code to test

Smooth human-computer interaction experience: the touch screen and the frame are integrated into the machine, and all consumables can be loaded online in real time


Model

BKI1100


Throughput

180T/H


Principle

Magnetic particle enzymatic chemiluminescence

Modes of Operation

Random, Batch and STAT

Separation method

Magnetic separation technology

Sample Positions

60 (Each position could be used as an emergency position)

Reagent Positions

30 pcs (2-8ºC refrigeration)

Incubation Positions

50


Time to 1st Result

18 Minutes

Sample Volume

10-200ul

Detection Precision

CV≤5%

Carry-over Rate

≤10-5

Correlation Coefficient

r≥0.99

Calibration Stability

28 Days

Calibration Type

6/7-point calibration

Software system

Windows 7/8/10, 32 or 64 bit
Bi-direction, support HL7 protocol
Intuitive User Interface, Intelligent Data Management

Interface

TCP/IP Network interface

Other Function

Liquid level detection
Anti-collision function
Sample barcode scanning
Reagent radio frequency identification
Intelligent alarm prompt

Power Supply

AC220V 50/60Hz (Standard) ;AC 110V, 60Hz (Optional)

External Size (W*D*H)

768*771*568mm


Net Weight

90Kg


Package Size (WD*H)

940*840*1105mm


Gross Weight

150kg



chemiluminescence Immunoassaychemiluminescent Immunoassay Analyzerautomatic Chemiluminescence Immunoassay Systemchemiluminescence Immunoassay